Abstract
A novel X-ray imaging system (NXRIS) and the design principles are given in this paper. Different from the existing digital X-ray imaging systems, the X-ray image intensifying system of NXRIS is a non-vacuum system composed of the intensifying screen and the brightness intensifier, and the brightness intensifier is named low light level image intensifier applied in military affairs. This structure makes NXRIS of big visual field (15 inch, even to larger) and low cost. When designing NXRIS, the spectral compatibility of the component devices and the relation between the visual field and the spatial resolution of the component devices are analyzed. The images produced by NXRIS are given and the image performance is good enough to be applied to security checking, non-destructive testing, and industry detection.
© 2005 Chinese Optics Letters
PDF Article
More Like This
Development of ZnO-based nanorod arrays as scintillator layer for ultrafast and high-spatial-resolution X-ray imaging system
Qianli Li, Xiaolin Liu, Mu Gu, Yahua Hu, Fengrui Li, Si Liu, Qiang Wu, Zhixiang Sun, Juannan Zhang, Shiming Huang, Zhijun Zhang, and Jingtai Zhao
Opt. Express 26(24) 31290-31298 (2018)
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Login to access Optica Member Subscription